Titanium Ti-6Al-4V ELI
The best strength-to-weight ratio among implantable metals. Its modulus, about half that of stainless steel, limits stress shielding; it is non-ferromagnetic and produces few MRI artifacts.
- Standards
- ISO 5832-3 · ASTM F136 · microstructure ISO 20160
- Composition
- Al 5.5–6.5% · V 3.5–4.5% · O ≤ 0.13% · Fe ≤ 0.25%
- Applications
- Spinal pedicle screws and rods, dental implants, osteosynthesis plates and screws, acetabular components.
- Density
- 4.43g/cm³
- Modulus E
- 110–114GPa
- UTS
- ≥ 860MPa
- YS 0.2%
- ≥ 795MPa
- Hardness
- ≈ 32–36 HRC
- Expansion α
- 8.610⁻⁶/K
- Note
- Condition: annealed. UTS ≥ 860 MPa · YS ≥ 795 MPa · El. ≥ 10% (ASTM F136 minimums)
Design and machining noteIn an implant, avoid direct contact with stainless steel or CoCr (galvanic corrosion, fretting). In machining, heat concentrates on the tool: moderate cutting speeds, abundant coolant.
At Precimade
- Processes
- Swiss-type turning, 5-axis milling, mill-turning
- Stocked bars
- Ø 2 to 65 mm, plates; Swiss-turning bars with UTS > 1,030 MPa
- Origin
- Bars of European or US origin, specified at order
- Finishing
- Bead blasting, vibratory finishing, color anodizing (subcontracted), UDI laser marking, cleaning